-
Notifications
You must be signed in to change notification settings - Fork 12
vertical_adjustment
Utility “Add Vertical Control” adds elevation information based on MOLA heights to the tie points measured in the Relative Orientation. The tie points are then flagged as Z-Control points in the ground point file (GPF).
The “Add Vertical Control” utility will auto-fill the input files based
on the <ProjectName>
listed. If the grayed out entries are
correct, simply check the boxes next to each field to confirm them.
Otherwise, press the buttons associated with each field to select the
correct files. The following steps detail the procedure.
-
From the SOCET SET menu bar, select
USGS Tools > Add Vertical Control
. -
If the project file listed in the
Project File
field is correct, check the box next to the Project File field to confirm it. Otherwise, press the “Project File…” button to bring up the list of projects and select the project from the list. -
The input ATF File should be
<ProjectName>_Relative.atf
. Either confirm the “ATF File” listed is correct by checking its check box, or press the “ATF File…” button to select<ProjectName>_Relative.atf.
(The Ground Point File (GPF) is also listed. It should be<ProjectName>.gpf
.) -
The input Grid File should be
<ProjectName>_mola.asc
located in theMOLA_DEM
folder. Either confirm theGrid File
listed is correct by checking its check box, or press theGrid File…
button to selectD:\\DATA\\<ProjectName>\\MOLA_DEM\\<ProjectName>_mola.asc
. -
The input Track File should be
<ProjectName>.tab
located in theMOLA_TRACKS
folder. Either confirm theTrack File
listed is correct by checking its check box, or press theTrack File…
button to selectD:\\DATA\\<ProjectName>\\MOLA\_TRACKS\\<ProjectName>.tab
. (Note: If a Track File does not exist, keep the Track File field blank and check the confirmation box.) -
Press
Execute
. (If the Execute button is not activated please check the confirmation checkboxes.) -
Close the window via the upper-right close button.
-
Open Windows Explorer. Navigate to
D:\\DATA\\<ProjectName>\\<ProjectName>_Original
, select all the files in the folder, and Copy them.Note: There are more than image support files in this folder, but there is no harm copying the extraneous files, and it is quickest to just select all the files.
-
Move up one folder so you are now in
D:\\DATA\\<ProjectName>
, and paste the files. -
In the pop-up window, check the box in the lower left corner to “do this for the next 3 conflicts”, then select “Copy and Replace”.
-
Close the Windows Explorer window.
-
From the SOCET SET menu bar, select
Preparation > Multi-Sensor Triangulation
. -
Select
File > Save Triangulation File As…
in the Automated Triangulation window.Note:
<ProjectName>\_Relative.atf
was automatically loaded. For the Vertical Adjustment to MOLA, no changes will be made to the Setup, however, we will provide a more meaningful name to the ATF file. -
In the pop-up window, enter
<ProjectName>_Rel_VertCtrl
, and pressOK
. -
Press
Solve
in the Automated Triangulation window. -
The
Simultaneous Solve
window will open, along with aBackup
pop-up window. PressYes
on the pop-up to overwrite back up files. -
In the
Simultaneous Solve
window, pressStart
to perform the bundle adjustment. -
Once the adjustment is completed, evaluate the errors in the adjustment.
An acceptable solution has (1) an Image (pixels) RMS of ~0.6 or less, and (2)no individual point measurement has an error greater than 2 pixels.
Press
Results
button in the Simultaneous Solve window to review error of individual point measurements. -
In the Vertical Adjustment procedure, all points are now Z-Control, so inspecting the top of the list will suffice.
-
If points have pixel errors greater than 2.0 pixels, SKIP TO: 18.4 Point Weights Refinement. Otherwise, continue to the next step.
-
If the image RMS is < ~0.6 pixels, and the maximum point measurement error is < 2.0 pixels, then Press
Close
on the Multi-Sensor Triangulation Residuals window. -
Press
Save
on the simultaneous solve window. -
Press
Yes
on the Done pop-up window. -
Press
Close
on the Simultaneous Solve window. -
Select
File > Exit
on the Automatic Triangulation window.Vertical Adjustment is now complete!
-
SKIP: 18.4 Point Weight Refinement and 18.5 Point Re-Measurement Process, GO TO: 18.6 Re-Load Images.
Continuing from 18.3 Bundle Adjustment…
The previous Relative Orientation adjustment should have met the criteria of the Image RMS less than ~0.6 pixels, and individual point errors less than 2.0 pixels. The only change since the Relative Orientation is that we added elevation estimates to the measured points. If the Image RMS < ~0.6 pixels; Point Errors < 2.0 pixels criteria is not met now, the cause is most likely that the weights (i.e., Accuracy Values) assigned to the elevation estimates by “Add Vertical Control” are too stringent. We will first evaluate and adjust Accuracy values. If the criteria of an acceptable solution is still not met after another bundle adjustment (Solve), then we will re-measure points.
-
From the
Multi-Sensor Triangulation Residuals
window, record (on a piece of paper) the Point ID’s of the points with > 2.0 pixel line and/or sample errors. Then close the Residuals window. -
Press
Close
on the Simultaneous Solve window. (Do not pressSave
.) -
Press
No
to Saving Files in the Done pop-up window. -
Press
Interactive Point Measurement
on the Automated Triangulation window. -
For each point recorded on your list:
a. Use the Scroll Bar to the right of the point list, and scroll to the point. Click on the point’s Point ID to select it.

b. IPM will move the stereo display to the selected point. Look at the measurement in stereo. If the point is on a slope, the Accuracy assigned to the Z coordinate may be too tight. We suggest you increase the Accuracy by approximately a factor of 2: double click in the Accuracy field for Z, delete the old value and type in the new value, then press the enter key.
-
When you are done updating Accuracy values, press
Save
thenClose
on the IPM window. -
Press
Solve
on the Automated Triangulation window. -
Press
Start
on the Simultaneous Solve window. -
Press
Residuals
on the Simultaneous Solve window and make sure no individual point has an error greater than 2 pixels. -
If the solution continues to have points with larger than 2 pixel errors, the problem may be a bad measurement(s), so GO TO 18.5 Point Re-Measurement Process. Otherwise, continue to the next step.
-
If the image RMS is < ~0.6 pixels, and the maximum point measurement error is < 2.0 pixels, we will exit Multi-Sensor Triangulation, and re-enter it in order to refresh the values stored in computer memory. First press
Close
on the Multi-Sensor Triangulation Residuals window. -
Press
Close
on the Simultaneous Solve window (Do Not PressSave
.) -
Press
No
to Saving Files in the Done pop-up window. -
Select
File > Exit
on the Automatic Triangulation window. -
Re-enter MST: From the SOCET SET menu bar, select
Preparation > Multi-Sensor Triangulation
. -
Press
Solve
on the Automatic Triangulation window. (The triangulation file,<ProjectName>_Rel_VertCtrl
should load automatically.) -
Press
Yes
on the pop-up to overwrite back up files. -
Press
Start
on the Simultaneous Solve window. -
If the Image (pixels) RMS is < ~0.6 pixels, Press
Save
on the Simultaneous Solve window. -
Press
Yes
on the Done pop-up window. -
Press
Close
on the Simultaneous Solve window. -
Select
File > Exit
on the Automatic Triangulation window.Vertical Adjustment is now complete!
-
SKIP: 18.5 Point Re-Measurement Process, GO TO: 18.6 Re-Load Images.
Continuing from 18.4 Point Weights Refinement…
-
On the
Multi-Sensor Triangulation Residuals
window, Left-Click on the Point ID(s) of the points to re-measure. (Hold the Control Key down to select multiple points.) Then pressRe-measure Point(s)…
. The re-measure point window will now open with the points selected. -
Left-Click on the point to re-measure. The View 1 (stereo display) window will display the current point measurement.
-
Un-check the Measured box for the Right image only, to remove parallax.
NOTE: Relocating the point is not advised because there is now an elevation estimate associated with the point.
-
Clear the parallax by moving the Right image only (i.e., put the dot on the ground).
-
Press
Sample
to collect the point measurement. -
Press
Save
write the measurement to disk. -
Repeat re-measurement process for remaining points in the list. (Go back to step 2.)
-
After points in the list are re-measured press
Close
. -
Press
Close
on the Multi-Sensor Triangulation Residuals window. -
Press
Start
on the Simultaneous Solve window. -
Press
Residuals
on the Simultaneous Solve window and make sure no individual point has an error greater than 2 pixels. -
If there are points with larger than 2 pixel errors, repeat the Point Re-Measurement Process starting from step 1.
-
If the image RMS is < ~0.6 pixels, and the maximum point measurement error is < 2.0 pixels, we will exit Multi-Sensor Triangulation, and re-enter it in order to refresh the values stored in computer memory. First press
Close
on the Multi-Sensor Triangulation Residuals window. -
Press
Close
on the Simultaneous Solve window (Do Not PressSave
.) -
Press
No
to Saving Files in the Done pop-up window. -
Select
File > Exit
on the Automatic Triangulation window. -
Re-enter MST: From the SOCET SET menu bar, select
Preparation > Multi-Sensor Triangulation
. -
Press
Solve
on the Automatic Triangulation window. (The triangulation file,<ProjectName>_Rel_VertCtrl
should load automatically.) -
Press
Yes
on the pop-up to overwrite back up files. -
Press
Start
on the Simultaneous Solve window. -
If the Image (pixels) RMS is < ~0.6 pixels, Press
Save
on the Simultaneous Solve window. -
Press
Yes
on the Done pop-up window. -
Press
Close
on the Simultaneous Solve window. -
Select
File > Exit
on the Automatic Triangulation window..
Vertical Adjustment to MOLA is now complete!
After the Vertical Adjustment to MOLA is complete, it is necessary to re-Load the images in order to view them with the results of the adjustment.
-
From the SOCET SET menu bar, select
File >
Load Images`. -
In the Image Loader window, select the Left and Right Image to display by clicking on the image id in the Left and Right panels. (Selected images will be highlighted.)
-
Under
View Control Panel
settings: Ensure thatView = 1
and thatLoad Point
is selected. -
Press
Load
.

-
Press
Close
.
At this point, it is prudent to backup the project data in order to have a re-entry point for the up-coming Absolute Orientation of the stereo images to MOLA.
-
From the SOCET SET menu bar, select
USGS Tools > Backup Orientation
. -
Make sure the current project name is listed in the Project field. If not, press “Project” to select the current project, then press
OK
. (Otherwise the backup will be made in, and for, the wrong project!) -
Replace
Backup_0
withRelative_VerticalControl
in theLocal
folder name field. The backup folder will be named<ProjectName>_Relative_VerticalControl
. -
Press
Backup
.