Software Manual and User Guide

This section contains an overview of the software design, instructions to build and install it, and a brief operator's guide.

Quickstart

The device ships with the software installed. The operator does not need to build or install anything.

At first startup from a powered-off state, the machine will self-calibrate by moving the piston to the extremeties. The LCD display will show CALIB in the status section (top-left) and will flash the green LED. The operator cannot interact with the device during this process.

After calibration, the LCD will display READY in the status section. The operator may change mode (Mo), breaths-per-minute (Hz), inhale-to-exhale ratio (IE), and volume limit (mL). These are selected by moving the dial at the top right, and edited by pressing the circular button at the bottom left. The display will show square brackets around the field to be edited, and at this time, movements of the dial will change the value. A change is committed by pressing the circular button, and aborted by pressing either of the triangular buttons.

To begin ventilation, press the right triangular button (which resembles a play button). The status will display RUNNING and the green LED will be solid. Parameters may be tuned while the machine is running.

To stop ventilation, press the left triangular button. The red LED will be solid and an alarm will sound. This is to ensure that an operator must always take notice when the machine is not running. At this point, any patient in the circuit should be removed and, if necessary, given manual ventilation support. The machine can be reset by pressing either of the triangular buttons. This will bring the piston back to the start position and the machine will enter the READY state again.

When the SIMV mode is selected, the machine will use the frequency, breaths-per-minute, and inhale-to-exhale ratios to determine a minimally acceptable breathing pattern. If the patient begins to inhale while the machine is not delivering a breath, it will begin to assist.

Build Instructions

A fully-automatic and reproducible build is expressed using nix. In the repository root directory, use nix-build to generate the binary along with a script to upload it to the microcontroller as follows:

$ nix-build default.nix --argstr port ttyACM0 --argstr usb false

where ttyACM0 is replaced with the actual serial interface to use. Change false to true if using the microcontroller's native USB port rather than its programming port. When this command finishes, the directory ./result will appear, containing the binary and an upload script:

$ ls ./result/
build.bin
upload
$ ./result/upload

Running upload will stop the microcontroller, erasing its current program, and upload the new one (build.bin). The microcontroller will restarts when this process has completed. The machine may be powered on while this process is carried out.

A precompiled binary is included in the root directory, called rideau.bin. This can be uploaded manually to the microcontroller using the BOSSA tool, available for all common end-user computing platforms.

System Description

The software system is factored into two components:

  • Low-level interfacing with sensors, switches, motor controls, human-input devices, etc.
  • High-level definition of the ventilation system itself.

The former is written in C, targeting the Arduino Mega, but in principle is appropriate for any board with similar or better capabilities in terms of I/O pins and microprocessor.

The second is expressed in Haskell, by way of the copilot suite of tools (BSD3 licensed). copilot was developed by Galois Inc. in partnership with NASA, to be applied in the runtime verification of unmanned aerial vehicle software[1]. A copilot specification is expressed in terms of pure functional streams of data--without destructive updates, pointers, or implicit type casts--and complies to C99 suitable for execution in low-memory, real-time environments, without loops or dynamic memory allocation.

In a software system where failure can lead to physical harm, high assurance is of utmost importance. The decision to use a domain-specific specification language, rather than a direct implementation in C, means it is easier to characterize, understand, and mitigate failure modes. As an added benefit, bug fixing, feature iteration, and general experimentation is much faster this way, due to the modularity and safety intrinsic to a pure functional language with a modern type system.

The ./spec directory contains the Haskell source files expressing the ventilator program abstract over hardware. The ./arduinobits/ directory contains the C source files expressing the interface between hardware and spec. At build time, the Haskell spec is used to generate the C source and header files ventilator.h and ventilator.c, which are used to compile and link the binary to be uploaded to the microcontroller.

1

Hazards and Limitations

Positive Characterization of Safe Operation

Rather than attempting to imagine every possible failure mode of the controller software, it would be more prudent to characterize exactly what is the condition of non-failure, i.e. what must the controller observe in order to judge that it is safe to move air to or from the patient. Whenever this condition is not met, the controller must raise visual and audible alarms and must not move any air until an operator (a medical professional) has diagnosed and resolved the problem.

Fractional Arithmetic and Use of Floating Point

Some software subsystems, notably kinematics and volume computation, pressure sensing, and flow sensing, are implemented using standard IEE754 floating point arithmetic. This is a safety hazard because errors may accumulate, leading to the controller delivering too little or too much air to the patient. It is essential that all such computations are meticulously audited to ensure that they fall within safe error bounds. Even if they were to be replaced with fixed point arithmetic, such an audit is required.

Unsafe casts from floating points to integers are present in the software and are at the moment not checked for unacceptable loss of accuracy.

API Description for Software Developers

There are generated docs for the high-level specification, to be viewed in any web browser.