This library is to communicate between an Arduino AVR processor and an S::CAN spectrometer via Modbus/RS485

License

Notifications You must be signed in to change notification settings

StroudCenter/S-CAN-Modbus

Repository files navigation

scan-Modbus

This library was created to communicate between an Arduino device and an s::can spectro::lyzer via Modbus/RS485. It is dependent on the SensorModbusMaster library. It is NOT intended to replace ana::pro, ana::lyte, or moni::tools as not all of the necessary settings for the spectro::lyzer can be controlled or set using the modbus commands. It is intended to allow spectro::lyzer data to be continuously uploaded to a network via wifi or cellular data streams without the very large power draw of a con::cube, con::stat, or con::lyte system.

In our arrangements, we are powering both the spectro::lyzer and the Arduino by way of an s::can con:nect. The con::nect itself is powered by a deep cycle marine battery with an attached solar panel. The con::nect is also necessary for the RS485 terminals. The RS-485 A and B terminals on the con::nect are wired to a RS-485 to TTL adapter board and from there to the serial port of the Arduino. When looking for an adapter board, be mindful of the TTL voltage levels! Because of the large quanitities of data going quickly between the Arduino and the spectro::lyzer, I very strongly reccommend using a hardware serial port on the Arduino. Hardware serial is also the only connection method which supports the spectro::lyzer's default of odd parity.

It is also possible to use this library to communicate with a con::cube, con::stat, or other s::can controller by way of ana::gate in Serial/RTU mode. The modbus map for ana::gate is very similar, but not identical, to the map for directly communicating wih the spectro::lyzer. The major difference is that when communicating with the spectro::lyzer directly, data is only accessible when the device is in logging mode but when communicating through ana::gate, the data is available when the device is in automatic or manual logging mode but not while in logging mode.

Please be cautious when using this library as much of the modbus mapping is not documented by s::can (including the registers containing the fingerprint data). Some of the documented functions actually do not work as described, either. The most completely mapping s::can provides of the modbus registers is in the manual for the con::cube, but this is still woefully incomplete. The remainder of the mapping I figured out myself by repeatedly scanning all of the registers on the spectro::lyzer and comparing the results with the data available in ana::pro. (I used the "scanRegisters" utility in the SensorModbusMaster library for this.) The FullSpecModbusMap.xlsx in this folder is my best guess of the full register mapping. There are still holes in my modbus map (which s::can has not been forthcoming about filling), so if you have any further information about the modbus mappings of the spectro::lyzer, PLEASE let me know. All issues and pull requests are welcome. NONE of this is in any way, shape, or form sanctioned by s::can! Please do not blame me if you "brick" your spec! Also, do not expect help or support from the s::can company in basically anything (related to modbus, broken instruments, anything). They will most likely tell you that your equipment must go back to Vienna to be fixed.


Specto::lyzer settings "safely" changable via modbus

Note: Only change one setting at a time. Ie., first change the serial baud rate, disconnect and reconnect the spec, then change the parity in a separate modbus call.

  • Sensor modbus address
  • Serial port settings (baudrate/parity)
  • Current device time
  • Initiating logging mode (when communicating directly with spectro::lyser, undocumented)
  • Setting the measurement interval (when communicating directly with spectro::lyser)
  • Opening and closing the cleaning valve (when communicating directly with spectro::lyser)
  • Setting the cleaning interval, duration, and wait period (when communicating directly with spectro::lyser)

Specto::lyzer settings NOT changable via modbus

  • Enabling and disabling sleep mode
  • Measurement parameters (flashrate, lamp voltage, etc)

Further specto::lyzer settings accessible via modbus

These settings are accessible, and theoretically changable, via modbus, but attempting to use modbus rather than ana::pro/moni::tools to change these settings is most likely going to lead to trouble.

  • Current scan::point name
  • Curernt parameter names, units, limits, and local calibrations (partially documented)
  • Current global calibration name (undocumented)
  • Current reference information, including all measurement parameters and absorbances of all stored references (varys by firmware, undocumented)

Data/results available via modbus

This data can be accessed, but not changed

  • Hardware meta-data (model, vendor, version, serial number)
  • Device status
  • Time most recent parameters were measured
  • Status and values of each parameter (single most recent measurement, only when in logging mode)
  • Fingerprint detector type, data source, and spectral path length (single most recent measurement, only when in logging mode, undocumented)
  • Time and status of fingerprint measurements (single most recent measurement, only when in logging mode, undocumented)
  • Full spectral values of 8 types of fingerprint data (single most recent measurement, only when in logging mode, undocumented)
    • 0 … Fingerprint [Abs/m]
    • 1 … Turbidity compensated fingerprint [Abs/m] (when communicating directly with spectro::lyser)
    • 2 … First derivative of fingerprint [Abs/m/m] (when communicating directly with spectro::lyser)
    • 3 … Difference between current print and print in memory [Abs/m] (when communicating directly with spectro::lyser)
    • 4 … Percent Transmission [%/cm2] (when communicating directly with spectro::lyser)
    • 5 … First derivative of turbidity compensated fingerprint [Abs/m/m] (when communicating directly with spectro::lyser)
    • 6 … Percent Transmission per 10cm2 [%/10cm2] (when communicating directly with spectro::lyser)
    • 7 … UNKNOWN (when communicating directly with spectro::lyser)

Data NOT accessible via modbus

  • Partial least squares components underlying global calibrations
  • Sample data used for local calibrations
  • "Historical" measurement data
    • only the single most recent data point is available

Available Examples and Utilities

These examples are in the "examples" folder:

  • "GetParameterValues" puts the spectro::lyzer into logging mode at 5-minute intervals and then prints the parameter values to the serial port every 5 minutes.
  • "SaveFingerprints" queries the spectro::lyzer and attempts to exactly re-create s::can's "par" and "fp" files on an SD card. It does not start the spectro::lyzer logging or make any attempt to change any of the spectro::lyzer's settings. It also does not put the Arduino to sleep between readings; even when fully active the Arduino only consumes ~1/10th of the power used by a sleeping spectro::lyzer.
  • "DisplayParamenter" is just like "SaveFingerprints", except that it also displays the parameter values to an I2C OLED display.

These utilities are also available in the "utils" folder:

  • "findSpec" searches for a response from the spec at all of the different baudrates, parities, and modbus addresses the spectro::lyzer typically supports. This could be really helpful if you do not know your spectro::lyzer's current settings. The default address seems to be 0x04, at 38400 baud, 8 data bits, odd parity, 1 stop bit. Not that this will only work when connecting to the spectro::lyzer with a hardware serial port.