Can't load Calibration #9

Open
huegene opened this issue May 31, 2020 · 4 comments
Open

Can't load Calibration #9

huegene opened this issue May 31, 2020 · 4 comments

Comments

@huegene
Copy link

huegene commented May 31, 2020

Loading any saved Calibration File ends up in an:
what(): matrix rank is not full! should be 4, is 3
Exception
(Current Appimage running on Debian)

@nanovna
Copy link
Collaborator

nanovna commented May 31, 2020

Can you post the saved calibration file?

@huegene
Copy link
Author

huegene commented Jun 2, 2020

One of the generated files would be
2400.zip

@nanovna
Copy link
Collaborator

nanovna commented Jun 4, 2020

the decimal points in the file are , instead of . (this should have been fixed in 20200507) what nanovna-qt version did you save the file with?

@huegene
Copy link
Author

huegene commented Jun 4, 2020

Just downloaded the most up to date Appimage, working now, Thanks a lot