-
Notifications
You must be signed in to change notification settings - Fork 1
/
scintillators.tex
194 lines (165 loc) · 17.6 KB
/
scintillators.tex
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
\section[Scintillation detectors]{Scintillation detectors \label{sec:scintillators}}
\subsection{Start counter \label{sec:st}}
The Start Counter (ST) detector, shown in Fig.~\ref{fig:st-overview-drawing},
surrounds the target
region and covers about 90\% of the solid angle for particles
originating from the center of the target. It is designed to operate
at tagged photon beam intensities of up to $10^8$ photons per second
in the coherent peak. It has a high degree of segmentation to limit
the the per-paddle rates. The time resolution must be sufficient to identify the electron RF bunch that generated the event
(see Section\,\ref{sec:ebeam}). The ST provides a timing signal that is relatively independent of particle type and trajectory (because of its proximity to the target), can provide particle identification via $dE/dx$, and can be used in the Level 1 trigger. Details of the design, construction and performance of the ST system can be found in
Ref.\,\cite{Pooser:2019rhu}.
\begin{figure}[!htb]
\centering
\includegraphics[width=1.0\columnwidth]{figures/start_counter_all.pdf}
\caption{The \gx{} Start Counter mounted to the liquid $\mathrm{H_2}$
target assembly. The beam goes from left to right down the central
axis.\label{fig:st-overview-drawing}}
\end{figure}
The ST consist of 30 scintillator paddles arranged in a cylinder for
most of its length with a ``nose'' section that bends towards the beam
line at the downstream end. EJ-200 scintillator from Eljen
Technology\footnote{Eljen Technology, https://eljentechnology.com/products/plastic-scintillators.} was
selected. EJ-200 has a decay time of 2.1~ns and an attenuation length
of 380~cm. Silicon photomultiplier (SiPM) detectors were used as light
sensors. These are not affected by the magnetic field produced by the
GlueX solenoid, an important feature since the start counter is located
near the geometric center of the field region. The SiPMs were placed
as close as possible to the upstream end of each scintillator element
to maximize light collection.
Each scintillator paddle started from stock 3~mm thick and 600 mm in
length. The paddles were bent at Eljen to create the nose section, and then machined at McNeal Enterprises Inc.\footnote{McNeal
Enterprises~Inc., http://www.mcnealplasticmachining.com} to their
final shape, including edges beveled at $6^\circ$ to minimize loss of
acceptance.
The scintillabor paddles are supported by a Rohacell closed-cell foam
structure. The Rohacell is 11~mm thick and is rigidly attached to an
aluminum support hub at its upstream end. The downstream support
extends partially into the nose section. The cylindrical length of the Rohacell is further reinforced with three layers of carbon fiber, each layer 650~$\mu$m thick. The assembly is made light-tight with a Tedlar wrapping. The Tedlar is attached to a plastic collar at the upstream end.
The scintillators are optically coupled to SiPM light sensors through a 250~$\mu$m air gap. Each paddle is read out with an array of four
magnetic-field-insensitive SiPMs (Hamamatsu S109031-050P multi-pixel photon counters), whose signals are summed.
The on-board electronics provides two signals per paddle, one delivered to a Flash ADC (FADC), the other to a 5$\times$~amplifier that is sent to a discriminator and then to an F1 TDC.
\subsection[Time-of-flight counters]{Time-of-flight counters \label{sec:tof}}
The Time-of-flight (TOF) detector is a wall of scintillators located about 5.5~m downstream from the target and covers
an angular region from 0.6$^{\circ}$ to 13$^{\circ}$ in polar angle. The detector has two planes of
scintillator paddles stacked in the horizontal and vertical direction, respectively. Most paddles are 252~cm long and 2.54~cm
thick with a width of 6~cm.
The scintillator material is EJ-200 from Eljen technology.
To accommodate the photon beam to pass through the central region,
an aperture of 12$\times$12\,cm$^2$ is kept
free of any detector material giving rise to four short paddle detectors with a length of 120~cm around the beam hole
in each detector plane. These paddles also have a width of 6~cm with a thickness of 2.54~cm. In order to keep the
count rate of the paddles well below 2~MHz the two inner most full-length paddles closest to the beam hole have a reduced width of 3~cm.
Light guides from UV transmitting plastic provide the coupling space between the scintillator and the PMT and allow the
magnetic shielding to protect the photo cathode by extending about 5~cm past the PMT entrance window. All paddles are wrapped
with a layer of a highly reflective material DF2000MA from 3M followed by a layer of strong black Tedlar film for light tightness.
The main purpose of the detector is to provide a fast timing for charged particles passing through the detector thereby providing information for particle identification. A detailed description of the TOF detector can be found in Ref.\,\cite{GlueXTOFNIM}.
The scintillator paddles are read out using
photomultiplier tubes (PMT) from Hamamatsu~\footnote{Hamamatsu Photonics, https://www.hamamatsu.com/us/en/index.html.}. Full-length paddles
have a PMT at both ends, while the short paddles have a single PMT
at the outer edge of the detector. These tubes of type H10534 have 10-stages and are complete assemblies with high voltage base, casing and $\mu$-metal shielding. Due to the significant stray field from the spectrometer solenoid magnet, additional external
shielding based on soft iron is necessary to protect the PMTs from the magnetic field.
\subsection{Electronics \label{sec:scelectronics}}
The high voltage (HV) to the TOF PMTs is provided by CAEN HV modules of type A1535SN initially controlled by a CAEN SY1527 main frame and
later upgraded to a SY4527.
The PMT output is connected to a splitter by a 55' long cables RG-58 coaxial cables. The signal is split by
a passive splitter into two equal-amplitude signals. One signals is directly connected to a flash ADC250
analog to digital converter (fADC)~\cite{Dong:2007}, while the second signal passes first through a leading edge discriminator (LED) and then used as an input to a high resolution TDC VX1290A from CAEN~\footnote{CAEN "https://www.caen.it/"}. The digitizer electronics (fADC250s and TDCs) are mounted in VXS crates as described in Section~\ref{sec:trig}.
The threshold of the leading edge discriminator is controlled for each channel separately and has an intrinsic
dead time of about 25~ns.
The sparcification threshold for the fADC250 is set to 120 (160) counts for the ST (TOF), with the nominal pedestal set at 100. The data from the fADC250 is provided by the FPGA algorithm and consists
of two words per channel with information about pedestal, signal amplitude, signal integral and timing.
The times of the ST system are registered using the JLab F1 TDCs, which have a nominal least count of 58~ps. In order to take advantage of the higher intrinsic resolution of the TOF counters, this system uses the VX1290A TDCs, which are multi-hit high-resolution TDCs with a buffer of up to 8 words per channel and a nominal least count of 25~ps. Since these TDCs provide the best time measurements in the GlueX detector, the timing of the accelerator RF signal is also
digitized using this electronics.
\subsection{Calibration and monitoring \label{sec:sccalib}}
The combined ST and TOF systems are used to determine the flight times of particles, the ST providing a precise start time in combination with the accelerator RF, and the TOF providing the stop time. Both systems may also be used to provide information on particle energy loss. Therefore, they must be
calibrated to determine corrections for the effects of
time-walk, light propagation time offsets, and light attenuation. The procedures are slightly different for the two detectors because of their different geometries, intrinsic resolutions and the advantages of the TOF system of having two adjacent perpendicular planes.
We start by describing the calibration procedures for the ST, which are performed for each scintillator separately.
For the time-walk correction, we use the fact that the detector signal is sent to both a Flash ADC and a TDC and the time from the the FADC is largely independent of pulse amplitude. This is used to measure the time walk seen by the TDC as a function of pulse amplitude as measured by the FADC; the resulting curve is fit to an empirical function for use in the correction. The propagation time is measured as a function of the hit position in a paddle as determined by well-reconstructed charged particle tracks. The propagation velocity is measured in three regions of the counter (``straight,'' ``bend,'' and ``nose'') and is not assumed to be a single value for all hits. The light attenuation is measured at several positions along the counter using charged particle tracks again. The energy per unit path length in the paddle as a function of distance from the SiPMs is fit to a modified exponential, with different parameters allowed for the straight section versus the nose with continuity enforced at the section boundary.
The calibration procedures for the TOF system take advantage of the fact that the detector consists of two
planes of narrow paddles oriented orthogonal to each other, which makes it possible to calibrate the full detector independent
of any other external detector information. The overlap region of two full-length paddles from the two planes define
a 6$\times$6~cm$^2$ area for most paddles with a few 3$\times$3~cm$^2$ areas close to the beam hole. The separation between the two detector planes is minimal as they are mounted on top of each other and as such are only separated by their wrapping
material. While the time-difference TD between the two ends of a paddle is related to the hit position along the paddle
the mean-time MT is related to the flight time of a particle from the vertex to the paddle. Therefor the MT for two overlapping
paddles must be the same when they are hit by the same particle passing through both of them while the hit position in the horizontal (x) and vertical (y) dimensions are defined by TD of the two paddles. This relationship results in an internally consistent calibration of all paddles with respect to every other paddle.
Prior to finding timing offsets, all times must be corrected for time-walk because of the use of LED discriminators, which
introduce a time shift that depends on the signal amplitude. The relation between time at threshold and signal amplitude is parameterized and used to correct for time slewing.
After all full-length paddles have been calibrated, they can be used themselves as references to
calibrate the remaining 8 short paddles that only have single-ended readout. Again we use the fact that any overlap region of two paddles from different
planes has the same particle flight time from the vertex. This coincidence produces peaks in the time difference distributions that can be used to determine the timing offsets of these single-ended readout paddles.
To test the calibration, we take tracks that are incident on a paddle in one plane and compute the time difference between the MT of that paddle and the MT of every other full-length paddle in the other plane. The resulting distribution of these differences is shown in Fig.~\ref{fig:mt_diff}. Assuming that all paddles have the same timing resolution, we can compute the
average time resolution to be $\sigma$ = 96~ps$=\frac{136}{\sqrt{2}}$~ps, assuming a Gaussian distribution.
\begin{figure}[tbp]
\begin{center}
\includegraphics[width=0.6\textwidth]{figures/mt_diff_fullTOF.pdf}
\caption{\label{fig:mt_diff} Mean time difference between one TOF long paddle of one plane with all other long paddles
of the other plane. (Color online)}
\end{center}
\end{figure}
\subsection{Performance \label{sec:scperformance}}
We begin by considering the time resolution of the ST. We note that the purpose of this system is to select the correct beam bunch that generated an interaction in the target, which will be used to determine the event start time using the accelerator RF time. Therefore, the ST resolution does not contribute to the resolution of the flight time as long it is sufficient to pick out the correct beam bunch with high probability.
The ST timing performance can be measured by comparing the event time in the target as measured by the start counter and the time derived from a signal from the CEBAF accelerator which is synchronized with the RF time structure of the machine. The start counter time must be corrected for the flight path of the charged particle emerging from the event and all instrumental corrections mentioned in the previous section are applied. Fig.~\ref{fig:st-time-resolution} shows the distribution of this time difference. The average time resolution is $\sigma$=234~ps, although it varies depending on the position of the hit along the counter.
%Table~\ref{table:st-time-resolution} gives the measured time resolution for the various sections as well as for all sections, with all paddles combined. Also shown is the fraction of tracks kept by a $\pm 1.0$~ns cut around the central value.
\begin{figure}[tbh]
\centering
\includegraphics[width=0.6\linewidth]{figures/st_tr_fit.pdf}
\caption{Time difference distribution between the vertex time computed from the start counter and the accelerator RF. The time from the RF does not contribute significantly to the width of the distribution. The vertical lines indicate the cuts used to identify a 250~MHz beam bunch.}
\label{fig:st-time-resolution}
\end{figure}
%\begin{table}[htbp]
% \centering
% \begin{tabular}{@{} l *4c @{}}
% \hline
% \multicolumn{1}{c}{\textbf{Section}} & \textbf{All} & %\textbf{Straight} & \textbf{Bend} & \textbf{Nose} \\
% \hline
% $\mathbf{FWHM}$ & 550~ps & 690~ps & 700~ps & 450~ps \\
% \textbf{Fraction} & 93\% & 92\% & 91\% & 94\% \\\hline
% \end{tabular}
% \caption{Average time resolutions (FWHM) and event fractions within a
% $\pm$ 1~ns window for all 30 ST sectors by independent geometrical regions.}
% \label{table:st-time-resolution}
%\end{table}
The ST can also be used to identify particles using its measurements of differential energy loss per unit path length ($dE/dx$). Fig.~\ref{fig:ST_dEdx_vs_p} shows $dE/dx$ vs. momentum
for a charged particles matched to the Start Counter. Protons can be
separated from pions up to 0.9 GeV/$c$ in momentum.
\begin{figure*}[!htb]
\centering
\includegraphics[width=0.6\textwidth]{figures/st_dedx_vs_p.pdf}
\caption{$dE/dx$ vs.\ $p$ for the Start Counter. The curved band
corresponds to protons while the horizontal band corresponds to
electrons, pions, and kaons. Pion/proton separation is achievable
for tracks with $p < 0.9$~GeV/$c$.}\label{fig:ST_dEdx_vs_p}
\end{figure*}
To investigate the performance of the TOF detector for its PID capability it is important to consider the relative number of
particle types within the event sample. We select events that have at least three fully reconstructed positively charged tracks with at least one of these tracks intersecting the TOF detector. We expect more pions than protons, and more protons than kaons. Looking at the distribution of velocity ($\beta$) of these tracks as a function of momentum it is easy to identify the bands from protons, kaons and pions (see Fig.~\ref{fig:betavsp}).
The distributions of $\beta$ at two specific track momenta, 2~GeV/c and 4~GeV/c (see Fig.~\ref{fig:betaproj}), are very illustrative of the PID capability of the TOF detector. At 2~GeV/c particle momentum the TOF detector provides about a 4$\sigma$ separation between
the pion/positron peak and the kaon peak. This is sufficient to identify tracks with a $\beta$ of 0.97 or lower as kaons with a very
high certainty. However, at a $\beta$ of 0.98 the probability of the track begin a kaon is less than 50\% mainly due to the fact
that the abundance of pions is close to an order of magnitude larger than kaons. The protons, on the other hand, are very well
separated from the other particle types and can be identified as such with high confidence over the full range in $\beta$.
At a track momentum of 4~GeV/c, this has changed and represents the limit at which the TOF can identify protons with high confidence. Again the separation between the large peak containing pions, kaons and positrons is separated from the proton
peak by about 4$\sigma$, while the relative abundance in this case is about a factor of 4. As a consequence, a 4~GeV/c momentum
track with a $\beta$ of 0.975 is most likely a proton with a small probability of being a pion. At a $\beta$ of 0.98 such
a track has a similar probability for being a proton or a pion.
\begin{figure}[tbp]
\begin{center}
\includegraphics[width=0.6\textwidth]{figures/beta_vs_p_positivetracks.pdf}
\caption{\label{fig:betavsp}$\beta$ of positive charged track vs track momentum. The color coding of the third dimension
is in logarithmic scale.(Color online)}
\end{center}
\end{figure}
\begin{figure}[tbp]
\begin{center}
\includegraphics[width=0.45\textwidth]{figures/TOF_postracks_2000mev.pdf}
\includegraphics[width=0.45\textwidth]{figures/TOF_postracks_4000mev.pdf}
\caption{\label{fig:betaproj}$\beta$ of positive charged track with 2~GeV/c momentum (left) and with 4~GeV/c (right).}
\end{center}
\end{figure}
\subsection{Summary \label{sec:scsummary}}
The TOF detector in the forward region of the GlueX spectrometer provides high resolution timing information that contributes
to the identification of the RF beam bucket of the beam photon that initiated the event. In combination with the charged
track reconstruction, TOF hits can be matched to such tracks to determine $\beta$ and give access to the particle
mass of the track. Protons can be identified with reasonable confidence up to momenta of 4~GeV/c while kaons can be
identified up to momenta of 2~GeV/c.