Analytical procedure for automated LC/MS/MS assay of digoxin and dihydrodigoxin in plasma
Modified from Yong Huang 6/16
Version 1.0 16 May 2018 ENB + JB
Reagents, consumables, and equipment
Standard
Company
Catolog #
Stock Concentration
Stock Diluent
Digoxin
Sigma
D6003
500µg/mL
DMF/MeOH
Oleandrin(IS)
Sigma
06069
500µg/mL
MeOH
Dihydrodigoxin
Custom synthesized
NA
500µg/mL
DMF/MeOH
Pump: Shimadzu LC-30 AD
Injector: Shimadzu SIL- 30AC set at 4°C
Detector: LC/MS/MS API Qtrap 6500
Column: ODS Hypersil 4.6 x 100 mm, 5 µm particle size (or equivalent)
Mobile phase: 65/35 (v/v) MeOH:water
Column Temperature: 50°C
Inection Volume: 10-20µL
Run Time 5.0min
Metabolite
Elution Time(min)
Digoxin
2.0
Dihydrodigoxin
2.1
Oleandrin(IS)
3.6
Sample Inlet: Turbo Spray IonDrive
Mass scanning mode: Negative MRM (Multiple Reaction Monitor)
Metabolite
Q1 (m/z)
Q3 (m/z)
Digoxin
779.4
475.2
Dihydrodigoxin
781.4
521.1
Oleandrin(IS)
575.2
531.1
Ionization: ESI/Negative Ionization
Parameter
Setting
Duration
5.0min
Cycle Time
0.615sec
Cycles
488
Scan Time
200 msec
Scan Type
Negative MRM
Resolution Q1
unit
Resolution Q3
unit
Intensity Threshold
0.00 cps
Smart Setting
off
Settling Time
0.00 msec
MR Pause
5.00 msec
MCA
No
GS1
55
GS2
55
CUR
20
Temperature
600°C
CAD
High (N2 = 99.999%)
Step Size
0.00 amu
DP
-40
EP
-10.00
CE Digoxin
-59
CE Dihydrodigoxin
-56
CE Oleandrin (IS)
-45
CXP
12
IS
4500
Weighted linear regression
1/x2
Standards are prepared by mixing 45µL H2 O with 5µL serum and the volumes below:
Standard
50% ACN (µL)
Dig Working Solution (ng/mL)
Working Solution (µL)
Stanrdard Curve Concentration (ng/ml)
00
20
0
0
0
0
20
0
0
0
1
18
2.5
2
0.1
2
15
2.5
5
0.25
3
10
2.5
10
0.5
4
0
2.5
20
1
5
18
50
2
2
6
15
50
5
5
7
10
50
10
10
8
0
50
20
20
QC samples are prepared according to the volumes below:
QC Sample
50% ACN (µL)
Dig Working solution (ng/mL)
Volume Spiked (µL)
Nominal Concentration)
Low
14
2.50
6
0.75
Medium
18
50
2
5.0
High
4
50
16
40
Extraction