As mentioned in this slack message, since a readout parameter change on 20250114 there is a degradation of a hot column in z1, and it significantly affects the overscan region of amp C. This led to a number of fibers in petal 1 to be flagged as "LOWEFFTIME" (desihub/desisurveyops#274).
To illustrate, below are morning darks before and after the readout change. The overscan region is on the right side of the red line. We will try a software fix by discarding the contaminated rows in the overscan image. The right panel shows another hot column on the same CCD which may be insufficiently masked at the moment.
The text was updated successfully, but these errors were encountered:
As mentioned in this slack message, since a readout parameter change on 20250114 there is a degradation of a hot column in z1, and it significantly affects the overscan region of amp C. This led to a number of fibers in petal 1 to be flagged as "LOWEFFTIME" (desihub/desisurveyops#274).
To illustrate, below are morning darks before and after the readout change. The overscan region is on the right side of the red line. We will try a software fix by discarding the contaminated rows in the overscan image. The right panel shows another hot column on the same CCD which may be insufficiently masked at the moment.