From bed7be3c9fb03492a889204e0d4addd37c82d321 Mon Sep 17 00:00:00 2001 From: Mehdi Zaidi <55298601+meh-di@users.noreply.github.com> Date: Wed, 6 Nov 2024 15:56:59 -0500 Subject: [PATCH] Add optimized parameters for 20uL tip + New Design P50 --- .../hardware_testing/gravimetric/liquid_class/defaults.py | 4 ++-- 1 file changed, 2 insertions(+), 2 deletions(-) diff --git a/hardware-testing/hardware_testing/gravimetric/liquid_class/defaults.py b/hardware-testing/hardware_testing/gravimetric/liquid_class/defaults.py index 4eaf1d15f1e..3a6f781ddf5 100644 --- a/hardware-testing/hardware_testing/gravimetric/liquid_class/defaults.py +++ b/hardware-testing/hardware_testing/gravimetric/liquid_class/defaults.py @@ -29,11 +29,11 @@ 1: DispenseSettings( # 1uL z_submerge_depth=_default_submerge_dispense_mm, plunger_acceleration=_default_accel_p50_ul_sec_sec, - plunger_flow_rate=57, # ul/sec + plunger_flow_rate=22, # ul/sec delay=_default_dispense_delay_seconds, z_retract_discontinuity=_default_retract_discontinuity, z_retract_height=_default_retract_mm, - blow_out_submerged=7, + blow_out_submerged=3.5, ), 10: DispenseSettings( # 10uL z_submerge_depth=_default_submerge_dispense_mm,